ANALISIS UJI KESELAMATAN LISTRIK DAN KALIBRASI CENTRIFUGE

Authors

  • Hidayah Rizka Universitas Sari Mutiara Indonesia
  • Yulizham Universitas Sari Mutiara Indonesia
  • Khairil Abdillah Universitas Sari Mutiara Indonesia
  • Sri Ulina Universitas Sari Mutiara Indonesia

DOI:

https://doi.org/10.51544/elektromedik.v6i2.3569

Keywords:

Testing, Environmental Conditions, Centrifuge, Safety, Calibration

Abstract

Health equipment is a supporting tool for medical service facilities. Room temperature is also very important in the placement of medical devices so that there are no errors in reading the equipment or damage to medical devices can occur. Calibration is an activity to determine the conventional truth of the value of the indication of measuring instruments and measuring materials by comparing them to traceable measuring standards to national and international standards for units of measure and/or international and certified reference materials. The method used in conducting electrical safety test analysis and calibration on the centrifuge is ECRI, IEC 60601-2.

The results of environmental measurements at temperature and humidity meet the requirements with a measured room temperature of 23°C and room humidity of 57.1%, the results of the electrical safety test measurement are declared safe and measured. Protective earth resistance is 0.432mΩ, equipment leakage current is 0.1µA and insulation resistance is 11.1MΩ , the results of the rotational speed accuracy measurement meet the tolerance limits of each value, for setting 2000RPM, 2500RPM, 3000RPM, 3500RPM, 4000RPM there is a deviation of 14.8 with an uncertainty value of 7.0 RPM, the results of the time accuracy measurement meet the tolerance limit with a setting value of 5 minutes (300 seconds) there is a deviation of 1.6 seconds with an uncertainty value of 2.55 seconds, for a full automatic Centrifuge with type HC-16D with Serial No. 2H01018 declared fit for use because it meets the tolerance limit.

Downloads

Download data is not yet available.

References

Andrian Ryan Pratama 2017, Alat Kalibrasi Centriduges Non Contact Menggunakan Microcontroller Atmega 8, skripsi, Universitas Muhamadiyah , Yogyakarta.

Burhanudin, M 2011. Dasar ketidakpastian pengukuran, jakarta : Salemba Medika.

Dewan Standarisasi Nasional/DNS, Pengukuran Kualitas Sistem Informasi Manajemen Kalibrasi Alat Medis (SIMKAM) di Rumah Sakit, Jurnal Kesehatan Masyarakat Indonesia.

Djonosaputro, B Darmawan., 1984, Teori Ketidakpastian Menggunakan SI , Penerbit ITB Bandung

ISO/IEC, 2005, Persyaratan Umum Kompetensi Laboratorium Pengujian dan Laboratorium Kalibrasi, Standar Nasional Indonesia.

KIM LIPI., 2016, Defenisi kalibrasi, pusat penelitian metrologi, Tangerang.

LIPI., 2016, ketidakpastian pengukuran, Pusat Penelitian Metrologi, Tangerang

Downloads

Published

2022-12-28