ANALISA PENYELESAIAN MASALAH DAN PERBAIKAN ALAT PATIENT MONITOR DI RSUP H. ADAM MALIK MEDAN
DOI:
https://doi.org/10.51544/elektromedik.v7i1.4496Keywords:
Patient monitor, Troubleshooting, RepairAbstract
Research on patient monitor tools has been carried out. The purpose of this study was to determine the problem of repairing the patient monitor tool. Corrective maintenance is maintenance action taken to overcome damage or congestion that occurs repeatedly, then take corrective action on the tool. Patient monitor is a tool that is used to monitor the patient's physiological condition. The purpose of this study is Troubleshooting (damage) patient monitor tools. Improvements to the patient monitor tool with the results of the study found that troubleshooting was seen through the results of less accurate blood pressure calculations so that it affected the diagnosis process in NIBP (Non Invasive Blood Pressure), Spo2 (Oxygen Pressure Saturation) patients. Incorrect reading of SpO2 results can be affected by sensor damage which when diagnosing patients, Ecg (Electrocardiograph) affects the results of diagnostics made by the instrument on patients, Misdiagnosis can occur due to incorrect placement of electrodes/tappers, damaged electrodes, Temperature (temperature) on the temperature sensor, namely damage to the temperature connector cable and the occurrence of undetected temperature results on the patient monitor tool. The research is that the patient monitor has been functioning properly after checking and repairing. Maintenance of the patient monitor must be routine and scheduled to prevent damage so that the age of use of tools and accessories is longer. Testing and calibration of patient monitor tools, especially on patient monitor parameters, must carry out routine checks on patient monitors, especially patient monitor accessories.
Downloads
References
B40/B20 Patient Monitor User's Guide.(2014). B40/B20 Patient Monitor User's Guide.diakses dari; https://image.tigermedical.com/Manuals/GEH2060600- 005-507513--20180315104424922.pdf, 21 November 2014.
Diakses dari; https://repository.usm.ac.id/files/skripsi/,2019. Ayada. (2012). How Equipment Works. Retrieved 06 2018, 24, diakses dari; http://www.howequipmentworks.com/pulse_oximeter/,British MedicalAssociation, 2005.
Fuad, Ahmad. 2014. Pasien Monitor, diakses dari: http://dumedpower.com/tag/pasienmonitor/, 11 November 2014.
Medical Systems.,Inc,.Information Technologies,West Tower Avenue,USA.diakse dari;https://medaval.ie/docs/manuals/GE-Dash-3000-5000-Manual.pdf/10 Mei 2005.
Milwaukee,operating manual book:PASIEN MONITOR GE DASH5000, Published by GE
Nanjing Kaihon.(2018).Pemeliharaan dan perbaikan pasien monitor.diakses dari ; http://m.id.fr-khmedical.com/,21 juni 2018.
Reyhan Abi Negoro. (2018).komparasi uji tekanan non invasive blood pressure (NIBP) Pada pasien monitor.(Electronic Thesis or Dissertation). Diakses dari ;https://localhost/setiadi,2018.
Rizaldi Ibrahim Nst .(2020) .Pasien Monitor.diakses dari ; /PATIENT-MONITOR- NEW-1-pdf/ 29Mei 2020.
Shabot M.M.(1994).Decision Support Systemsin Critical Care,Springer- Verlag,Boston
Downloads
Published
Issue
Section
License
Copyright (c) 2023 Fitria Priyulida, Kesaktian Manurung, Fazri Maulana, Riris Malau
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.